Deck 6: Quality Assurance and Calibration Methods

ملء الشاشة (f)
exit full mode
سؤال
The percent by mass octane is quantified for an organic sample via GC using decane as the internal standard.A GC vial containing 0.5023 grams organic sample and 0.2384 grams octane is prepared and analyzed,returning peak areas of 119,993 for octane and 83755 for decane.If a response factor of 2.50 from previous work analyzing octane with a decane internal standard is used,what is the percent by mass octane in the organic sample?
استخدم زر المسافة أو
up arrow
down arrow
لقلب البطاقة.
سؤال
There are several types of precision involved with method validation.

A)instrument precision
B)final precision
C)intermediate precision
D)interlaboratory precision
E)intra-assay precision
سؤال
A____________________ is a sample that has been exposed to the site of sampling,containing all components except analyte,and it is taken through all steps of the analytical procedure.

A)method blank
B)reagent blank
C)field blank
D)spike blank
E)matrix blank
سؤال
A(n)________________________________________is a known amount of compound,different from the analyte,that is added to the unknown.

A)standard addition
B)internal standard
C)external standard
D)self-contained standard
E)known standard
سؤال
A set of five solutions are prepared by delivering 10 mL of unknown sample and increasing volumes of standard (0,5,10,15,and 20 mL)into 50-mL volumetric flasks.Analytical reagents are then added and diluted to volume.The x-intercept is −3.90 mM.What is the unknown concentration?

A)39.0 mM
B)156 mM
C)19.5 mM
D)0)390 mM
E)195 mM
سؤال
Which are TRUE statements regarding internal standards?
I The response factor is calculated by using a mixture containing known amounts of analyte and internal standard and by using detector response to the analyte and the internal standard.
II Internal standards are useful where instrument response or quantity of sample analyzed varies over time.
III Detector response is inversely proportional to concentration.
IV Detector response to analyte and internal standard is assumed to be constant over a range of concentrations.

A)I,II,and III
B)II,III,and IV
C)I and II
D)I and III
E)I,II,and IV
سؤال
The____________________ state(s)what steps will be taken and how they will be carried out.

A)assessment
B)calibration check
C)control charts
D)specifications
E)standard operating procedure
سؤال
Which of the following is(are)not a way to demonstrate the accuracy of a method?
I Determine the limit of detection for the method.
II Analysis of a certified reference material in a matrix similar to the sample.
III Compare the results between two or more different analytical methods.
IV Analyze a blank sample with a known amount of analyte in the same matrix as the sample.

A)Only IV
B)II and III
C)I and III
D)Only III
E)Only I
سؤال
____________________ is the concentration range over which there is measurable instrument response.

A)Accuracy range
B)Dynamic range
C)Linear range
D)Detection range
E)Range
سؤال
A____________________ is an analytical result that indicates the sample concentration is below a stated limit when the concentration is actually above the limit.

A)positive result
B)false negative
C)negative result
D)false positive
E)negative
سؤال
A solution is 2.0% by mass compound A and 3.5% by mass compound I.Separation via GC returns peak areas of 38234 for compound A and 23880 for compound I.Calculate the response factor.

A)0)664
B)2)03
C)1)50
D)0)491
E)1)17
سؤال
The detection limit for caffeine analysis must be calculated.Eight absorbance readings from a caffeine sample close to the estimated detection limit are 0.1239,0.1098,0.1182,0.1150,0.1219,0.1251,0.1199,and 0.1099.The average absorbance for eight blanks is 0.0182.The slope of a calibration curve is 0.59 /mM.What is the detection limit?
سؤال
Which statement regarding standard addition to a single solution is FALSE? A Analyte signal is measured after each standard addition.

A)The volume standard added must be kept small to minimize altering the matrix.
B)Detector response is corrected to take in account dilution before being plotted on the y-axis.
C)The absolute value of the y-intercept for the best fit line is the unknown analyte concentration.
D)Volume diluted standard is plotted on the x-axis.
سؤال
There are many studies required to validate an analytical method.Which of the studies below is not required for method validation?

A)method specificity
B)linearity
C)range
D)calibration
E)limit of detection
سؤال
A water sample collected downstream from a gold-mining operation returns a cyanide concentration of 6.18 μ\mu g/L CN.A 10-mL aliquot is spiked with 100 μ\mu L of 500.0 μ\mu g/L sodium cyanide and analyzed.Analysis returns a cyanide concentration of 10.5 μ\mu g/L CN.What is the percent recovery for the spike?
سؤال
Which statement regarding the multiple solutions with constant volume standard addition method is FALSE?

A)Equal volumes of unknown are pipetted into a set of volumetric flasks with identical volume.
B)Equal volumes of standard are pipetted into each volumetric flask.
C)Analytical reagents are added to each volumetric flask and diluted to volume.
D)Detector response versus diluted standard is plotted.
E)The absolute value of the x-intercept for the best-fit line is the diluted unknown concentration.
سؤال
A chemist is tasked with determining the concentration of nickel for a water sample drawn from a well next to a nickel mine.To analyze the sample using atomic absorbance spectroscopy (AAS),she prepares six samples (0-5)by adding 10.00 mL well water to a 50.00-mL volumetric flask,then adding an increasing volumes of 25.00-ppm nickel solution,diluting the resulting solution to 50.00 mL with 0.1 M HNO3.Following AAS analysis,a graph of absorbance versus ppm nickel standard gave a best-fit line y = 0.0428x + 0.0487.Calculate the ppm nickel in the well water sample.
سؤال
A____________________ is a sample containing all components except analyte,and it is taken through all steps of the analytical procedure.

A)method blank
B)reagent blank
C)field blank
D)spike blank
E)matrix blank
سؤال
The response factor,F,for the analysis of pentane with the internal standard decane is 3.5.Which statements below are INCORRECT for the analysis?
I The pentane concentration is 3.5 times greater than the decane concentration.
II The decane concentration is 3.5 times greater than the pentane concentration.
III Detector response to pentane is 3.5 times greater than detector response to decane.
IV Detector response to decane is 3.5 times greater than detector response to pentane.

A)I and III
B)II and III
C)II and IV
D)I and IV
E)None of these is incorrect.
سؤال
The____________________ is everything in the sample,except analyte.

A)blind sample
B)performance test sample
C)spike
D)matrix
E)assessment
فتح الحزمة
قم بالتسجيل لفتح البطاقات في هذه المجموعة!
Unlock Deck
Unlock Deck
1/20
auto play flashcards
العب
simple tutorial
ملء الشاشة (f)
exit full mode
Deck 6: Quality Assurance and Calibration Methods
1
The percent by mass octane is quantified for an organic sample via GC using decane as the internal standard.A GC vial containing 0.5023 grams organic sample and 0.2384 grams octane is prepared and analyzed,returning peak areas of 119,993 for octane and 83755 for decane.If a response factor of 2.50 from previous work analyzing octane with a decane internal standard is used,what is the percent by mass octane in the organic sample?
27.3% octane
2
There are several types of precision involved with method validation.

A)instrument precision
B)final precision
C)intermediate precision
D)interlaboratory precision
E)intra-assay precision
final precision
3
A____________________ is a sample that has been exposed to the site of sampling,containing all components except analyte,and it is taken through all steps of the analytical procedure.

A)method blank
B)reagent blank
C)field blank
D)spike blank
E)matrix blank
field blank
4
A(n)________________________________________is a known amount of compound,different from the analyte,that is added to the unknown.

A)standard addition
B)internal standard
C)external standard
D)self-contained standard
E)known standard
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 20 في هذه المجموعة.
فتح الحزمة
k this deck
5
A set of five solutions are prepared by delivering 10 mL of unknown sample and increasing volumes of standard (0,5,10,15,and 20 mL)into 50-mL volumetric flasks.Analytical reagents are then added and diluted to volume.The x-intercept is −3.90 mM.What is the unknown concentration?

A)39.0 mM
B)156 mM
C)19.5 mM
D)0)390 mM
E)195 mM
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 20 في هذه المجموعة.
فتح الحزمة
k this deck
6
Which are TRUE statements regarding internal standards?
I The response factor is calculated by using a mixture containing known amounts of analyte and internal standard and by using detector response to the analyte and the internal standard.
II Internal standards are useful where instrument response or quantity of sample analyzed varies over time.
III Detector response is inversely proportional to concentration.
IV Detector response to analyte and internal standard is assumed to be constant over a range of concentrations.

A)I,II,and III
B)II,III,and IV
C)I and II
D)I and III
E)I,II,and IV
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 20 في هذه المجموعة.
فتح الحزمة
k this deck
7
The____________________ state(s)what steps will be taken and how they will be carried out.

A)assessment
B)calibration check
C)control charts
D)specifications
E)standard operating procedure
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 20 في هذه المجموعة.
فتح الحزمة
k this deck
8
Which of the following is(are)not a way to demonstrate the accuracy of a method?
I Determine the limit of detection for the method.
II Analysis of a certified reference material in a matrix similar to the sample.
III Compare the results between two or more different analytical methods.
IV Analyze a blank sample with a known amount of analyte in the same matrix as the sample.

A)Only IV
B)II and III
C)I and III
D)Only III
E)Only I
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 20 في هذه المجموعة.
فتح الحزمة
k this deck
9
____________________ is the concentration range over which there is measurable instrument response.

A)Accuracy range
B)Dynamic range
C)Linear range
D)Detection range
E)Range
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 20 في هذه المجموعة.
فتح الحزمة
k this deck
10
A____________________ is an analytical result that indicates the sample concentration is below a stated limit when the concentration is actually above the limit.

A)positive result
B)false negative
C)negative result
D)false positive
E)negative
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 20 في هذه المجموعة.
فتح الحزمة
k this deck
11
A solution is 2.0% by mass compound A and 3.5% by mass compound I.Separation via GC returns peak areas of 38234 for compound A and 23880 for compound I.Calculate the response factor.

A)0)664
B)2)03
C)1)50
D)0)491
E)1)17
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 20 في هذه المجموعة.
فتح الحزمة
k this deck
12
The detection limit for caffeine analysis must be calculated.Eight absorbance readings from a caffeine sample close to the estimated detection limit are 0.1239,0.1098,0.1182,0.1150,0.1219,0.1251,0.1199,and 0.1099.The average absorbance for eight blanks is 0.0182.The slope of a calibration curve is 0.59 /mM.What is the detection limit?
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 20 في هذه المجموعة.
فتح الحزمة
k this deck
13
Which statement regarding standard addition to a single solution is FALSE? A Analyte signal is measured after each standard addition.

A)The volume standard added must be kept small to minimize altering the matrix.
B)Detector response is corrected to take in account dilution before being plotted on the y-axis.
C)The absolute value of the y-intercept for the best fit line is the unknown analyte concentration.
D)Volume diluted standard is plotted on the x-axis.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 20 في هذه المجموعة.
فتح الحزمة
k this deck
14
There are many studies required to validate an analytical method.Which of the studies below is not required for method validation?

A)method specificity
B)linearity
C)range
D)calibration
E)limit of detection
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 20 في هذه المجموعة.
فتح الحزمة
k this deck
15
A water sample collected downstream from a gold-mining operation returns a cyanide concentration of 6.18 μ\mu g/L CN.A 10-mL aliquot is spiked with 100 μ\mu L of 500.0 μ\mu g/L sodium cyanide and analyzed.Analysis returns a cyanide concentration of 10.5 μ\mu g/L CN.What is the percent recovery for the spike?
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 20 في هذه المجموعة.
فتح الحزمة
k this deck
16
Which statement regarding the multiple solutions with constant volume standard addition method is FALSE?

A)Equal volumes of unknown are pipetted into a set of volumetric flasks with identical volume.
B)Equal volumes of standard are pipetted into each volumetric flask.
C)Analytical reagents are added to each volumetric flask and diluted to volume.
D)Detector response versus diluted standard is plotted.
E)The absolute value of the x-intercept for the best-fit line is the diluted unknown concentration.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 20 في هذه المجموعة.
فتح الحزمة
k this deck
17
A chemist is tasked with determining the concentration of nickel for a water sample drawn from a well next to a nickel mine.To analyze the sample using atomic absorbance spectroscopy (AAS),she prepares six samples (0-5)by adding 10.00 mL well water to a 50.00-mL volumetric flask,then adding an increasing volumes of 25.00-ppm nickel solution,diluting the resulting solution to 50.00 mL with 0.1 M HNO3.Following AAS analysis,a graph of absorbance versus ppm nickel standard gave a best-fit line y = 0.0428x + 0.0487.Calculate the ppm nickel in the well water sample.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 20 في هذه المجموعة.
فتح الحزمة
k this deck
18
A____________________ is a sample containing all components except analyte,and it is taken through all steps of the analytical procedure.

A)method blank
B)reagent blank
C)field blank
D)spike blank
E)matrix blank
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 20 في هذه المجموعة.
فتح الحزمة
k this deck
19
The response factor,F,for the analysis of pentane with the internal standard decane is 3.5.Which statements below are INCORRECT for the analysis?
I The pentane concentration is 3.5 times greater than the decane concentration.
II The decane concentration is 3.5 times greater than the pentane concentration.
III Detector response to pentane is 3.5 times greater than detector response to decane.
IV Detector response to decane is 3.5 times greater than detector response to pentane.

A)I and III
B)II and III
C)II and IV
D)I and IV
E)None of these is incorrect.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 20 في هذه المجموعة.
فتح الحزمة
k this deck
20
The____________________ is everything in the sample,except analyte.

A)blind sample
B)performance test sample
C)spike
D)matrix
E)assessment