Deck 12: Quality Control in Computed Tomography

ملء الشاشة (f)
exit full mode
سؤال
The minimum distance between two objects that allows them to be seen as separate and distinct would define what CT imaging parameter?

A)Low-contrast resolution
B)Uniformity
C)High-contrast resolution
D)Image noise
استخدم زر المسافة أو
up arrow
down arrow
لقلب البطاقة.
سؤال
The laser light field and radiation field should coincide to within ____ mm.

A)0.5
B)1.0
C)2.0
D)5.0
سؤال
For ACR CT accreditation, the Water CT Number and Standard Deviation (Noise) should be evaluated by the CT technologist

A)quarterly.
B)monthly.
C)weekly.
D)daily.
سؤال
A visual inspection of a CT unit should be evaluated at least _____ for ACR accreditation programs.

A)daily.
B)weekly.
C)monthly.
D)semi-annually.
سؤال
The CTDIᵥₒₗ pass/fail criterion for an adult abdomen is ____ mGy.

A)20
B)40
C)30
D)80
سؤال
The uniformity of the CT number of water should not vary by more than ____ HU from the center of the phantom to the periphery.

A)2
B)3
C)5
D)10
سؤال
The CT number of water should not deviate by more than ____ CT numbers from 0.

A)2
B)3
C)5
D)10
سؤال
Section thickness should not vary by more than ____ mm for sections greater than 5 mm.

A)±0.5
B)±1.0
C)±2.0
D)±5.0
سؤال
For a slice thickness of 5 mm or less, the slice thickness should not vary by more than ____ mm.

A)0.5
B)1.0
C)2.0
D)3.0
سؤال
Patient dose values for CT scans should remain within ____ of the manufacturer's specifications.

A)±2%
B)±5%
C)±10%
D)±20%
فتح الحزمة
قم بالتسجيل لفتح البطاقات في هذه المجموعة!
Unlock Deck
Unlock Deck
1/10
auto play flashcards
العب
simple tutorial
ملء الشاشة (f)
exit full mode
Deck 12: Quality Control in Computed Tomography
1
The minimum distance between two objects that allows them to be seen as separate and distinct would define what CT imaging parameter?

A)Low-contrast resolution
B)Uniformity
C)High-contrast resolution
D)Image noise
C
2
The laser light field and radiation field should coincide to within ____ mm.

A)0.5
B)1.0
C)2.0
D)5.0
C
3
For ACR CT accreditation, the Water CT Number and Standard Deviation (Noise) should be evaluated by the CT technologist

A)quarterly.
B)monthly.
C)weekly.
D)daily.
D
4
A visual inspection of a CT unit should be evaluated at least _____ for ACR accreditation programs.

A)daily.
B)weekly.
C)monthly.
D)semi-annually.
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
5
The CTDIᵥₒₗ pass/fail criterion for an adult abdomen is ____ mGy.

A)20
B)40
C)30
D)80
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
6
The uniformity of the CT number of water should not vary by more than ____ HU from the center of the phantom to the periphery.

A)2
B)3
C)5
D)10
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
7
The CT number of water should not deviate by more than ____ CT numbers from 0.

A)2
B)3
C)5
D)10
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
8
Section thickness should not vary by more than ____ mm for sections greater than 5 mm.

A)±0.5
B)±1.0
C)±2.0
D)±5.0
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
9
For a slice thickness of 5 mm or less, the slice thickness should not vary by more than ____ mm.

A)0.5
B)1.0
C)2.0
D)3.0
فتح الحزمة
افتح القفل للوصول البطاقات البالغ عددها 10 في هذه المجموعة.
فتح الحزمة
k this deck
10
Patient dose values for CT scans should remain within ____ of the manufacturer's specifications.

A)±2%
B)±5%
C)±10%
D)±20%